FDA recall record

K2M recall — FDA Z-2017-2020

The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking…

Class IIclassification
Z-2017-2020official record ID
May 27, 2020report date

Recall record at a glance

Official recordZ-2017-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmK2M, Inc.
Report dateMay 27, 2020
Recall initiation dateApr 13, 2020
QuantityN/A
LocationLeesburg, VA, United States

Official product description

K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90135, UDI # 10888857108486, Size 24x4mm, 18 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.

Why the product was recalled

The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.

Distribution information

Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 6, 2022

K2M Yukon Straight Rod 4.0 x120mm recall — FDA Z-1307-2022

Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or…

Class IIFDA: OngoingK2M, Inc.