Recall record at a glance
| Official record | Z-2084-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Integra LifeSciences Corp. |
| Report date | Jul 6, 2016 |
| Recall initiation date | May 20, 2016 |
| Quantity | 182 units |
| Location | Plainsboro, NJ, United States |
Official product description
K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws
Why the product was recalled
The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.
Distribution information
Nationwide Distribution to OK, CA, TX, WA, GA, NY, MS, NY, PA
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.