FDA recall record

Biomet JUGGERKNOT IN¿GUIDE PUNCH, 2.9MM recall — FDA Z-2399-2012

Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a…

Class IIclassification
Z-2399-2012official record ID
Sep 26, 2012report date

Recall record at a glance

Official recordZ-2399-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBiomet, Inc.
Report dateSep 26, 2012
Recall initiation dateAug 14, 2012
Quantity62
LocationWarsaw, IN, United States

Official product description

JUGGERKNOT IN¿GUIDE PUNCH, 2.9MM. The JuggerKnot 2.9mm In-Guide Punch is intended to be used to prepare a 2.9mm hole into bone in which the JuggerKnot 2.9mm Suture Anchor is to be placed. The JuggerKnot 2.9mm Suture Anchor is intended to be used to attach soft tissue to bone.

Why the product was recalled

Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft.

Distribution information

Nationwide Distribution-including the states of WA, IN, AK, PA, GA, SD, KS, NY, VA, AR, AL, OH, WI, MA, and CA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.