FDA recall record

Stanmore Implants Worldwide JTS External Drive Unit JTS Distal Femoral Extendible Implant recall — FDA Z-3104-2017

Three (3) complaints were identified in which the potential for utilization of the JTS Grower Drive in the incorrect orientation in relation to the limb…

Class IIclassification
Z-3104-2017official record ID
Sep 13, 2017report date

Recall record at a glance

Official recordZ-3104-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmStanmore Implants Worldwide Ltd.
Report dateSep 13, 2017
Recall initiation dateJul 4, 2017
Quantity15 units
LocationBorehamwood, N/A, United Kingdom

Official product description

JTS External Drive Unit JTS Distal Femoral Extendible Implant Femur is intended for the replacement of diseased or deficient bone in the distal femur.

Why the product was recalled

Three (3) complaints were identified in which the potential for utilization of the JTS Grower Drive in the incorrect orientation in relation to the limb entry point of the device, together with the potential to select incorrect settings for the extension procedure. As a result, the prostheses in such cases failed to extend as intended.

Distribution information

US Distribution to the states of : CA, FL, GA, KS, MI, NJ, OH, PA, SC, and TX.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.