FDA recall record

Smith & Nephew Journey BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210 recall — FDA Z-0001-2019

The data in the National Joint Registry of England, Wales and Northern Ireland (NJREWNI) and Australian Orthopaedic Association National Joint Replacement…

Class IIclassification
Z-0001-2019official record ID
Oct 10, 2018report date

Recall record at a glance

Official recordZ-0001-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSmith & Nephew, Inc.
Report dateOct 10, 2018
Recall initiation dateJun 13, 2018
Quantity10,792 units
LocationMemphis, TN, United States

Official product description

Journey BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 74021222, 74021223, 74021224, 74021225, 74021226, 74021227, 74021228, 74021229

Why the product was recalled

The data in the National Joint Registry of England, Wales and Northern Ireland (NJREWNI) and Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) indicate that the first generation JOURNEY BCS Knee System has a revision rate over 1.5 times the primary total knee arthroplasty device class average revision rates reported in those registries.

Distribution information

Worldwide distribution, including US nationwide and Puerto Rico, Austria, United Arab Emirates, Australia, Belgium, Canada, China, Switzerland, Germany, Denmark, Finland, France, Great Britain, India, Spain, Italy, Netherlands, Portugal, Sweden, Venezuela, South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records