Recall record at a glance
| Official record | Z-1511-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Ohmeda Medical |
| Report date | Mar 29, 2017 |
| Recall initiation date | Nov 18, 2016 |
| Quantity | 5,993 (5,080 US; 913 OUS) |
| Location | Laurel, MD, United States |
Official product description
IWS Infant Warmer System-Infant radiant warmers provide infrared heat, in a controlled manner, to neonates who are unable to thermoregulate based on their own physiology or necessitate external heat to smoothen the transition from the uterus to the external environment. They fulfill the same purpose of neonatal incubators but they may be preferred over incubators when total access to the patient is needed or desirable (e.g., surgical procedures, extracorporeal membrane oxygenation, frequent resuscitation, etc.). Most infant radiant warmers can be used in two operating modes: 1. Manual: The clinician sets the appropriate heater output for maintaining the desired patient temperature. The heater output is initially selected based on the clinician s training and experience and then is adjusted based on the patients needs and clinical status. 2. Servo: The clinician sets the desired patient temperature. A skin temperature probe senses the patient temperature and feeds this information to the controller of the infant radiant warmer. The controller then adjusts the heater output to maintain the patient temperature at the set value. These adjustments to the heater output are made in such a way to gradually change the patient s temperature while minimizing overshooting and patient stress. Infant radiant warmers have alarms to alert clinicians when certain patient or equipment conditions occur, such as a malfunction, or an excessive departure of the patient s temperature from the set value. Infant radiant warmers may incorporate other features, such as phototherapy, observation light, tilting of the bed, and data output to remote monitors or nurse call systems
Why the product was recalled
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Distribution information
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.