Recall record at a glance
| Official record | Z-0154-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | ICU Medical, Inc. |
| Report date | Oct 29, 2025 |
| Recall initiation date | Oct 2, 2025 |
| Quantity | 23203 |
| Location | San Clemente, CA, United States |
Official product description
IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014
Why the product was recalled
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
Distribution information
Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.