Recall record at a glance
| Official record | Z-2136-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Beckman Coulter Inc. |
| Report date | Jun 3, 2020 |
| Recall initiation date | Jan 29, 2020 |
| Quantity | 2738 vials |
| Location | Brea, CA, United States |
Official product description
ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.
Why the product was recalled
Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.
Distribution information
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries of CANADA, JAPAN, LEBANON, SAUDI ARABIA, TAIWAN.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.