Recall record at a glance
| Official record | Z-0161-2027 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Beckman Coulter, Inc. |
| Report date | Oct 21, 2026 |
| Recall initiation date | Aug 12, 2026 |
| Location | Miami, FL, United States |
Official product description
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
Why the product was recalled
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
Distribution information
Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: not stated.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.