FDA recall record

Bard Access Systems Intraosseous needle and driver recall — FDA Z-1397-2022

BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of…

Class Iclassification
Z-1397-2022official record ID
Aug 3, 2022report date

Recall record at a glance

Official recordZ-1397-2022
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmBard Access Systems, Inc.
Report dateAug 3, 2022
Recall initiation dateJun 20, 2022
Quantity40,734
LocationSalt Lake City, UT, United States

Official product description

Intraosseous needle and driver. Catalog/Device Name: D015151NK/BD Needle Kit for Powered Driver 15mm x 15Ga, D015151MK/ BD Manual Driver Needle Kit 15mm x 15Ga, D015251NK/ BD Needle Kit for Powered Driver 25mm x 15Ga, D015251MK/ BD Manual Driver Needle Kit 25mm x 15Ga, D015351NK/ BD Needle Kit for Powered Driver 35mm x 15Ga, D015351MK/ BD Manual Driver Needle Kit 35mm x 15Ga, D015451NK/ BD Needle Kit for Powered Driver 45mm x 15Ga, D015451MK/ BD Manual Driver Needle Kit 45mm x 15Ga, D015551NK/ BD Needle Kit for Powered Driver 55mm x 15Ga, D015551MK/ BD Manual Driver Needle Kit 55mm x 15G, D001001/ BD Intraosseous Powered Driver (drill)

Why the product was recalled

BD Intraosseous Needle Kits may exhibit: 1) Increased force required to remove the stylet from intraosseous needle during placement may cause removal of the entire needle assembly and loss of intraosseous access; 2) Stylet safety mechanism may not engage as the stylet is removed; 3) Metal discs in the powered driver may stick rendering the driver unusable.

Distribution information

Distribution US nationwide and Canada.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.