FDA recall record

Edwards Lifesciences IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) recall — FDA Z-1705-2019

Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the…

Class Iclassification
Z-1705-2019official record ID
Jun 19, 2019report date

Recall record at a glance

Official recordZ-1705-2019
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmEdwards Lifesciences, LLC
Report dateJun 19, 2019
Recall initiation dateMay 14, 2019
Quantity6552
LocationIrvine, CA, United States

Official product description

IntraClude Intra-Aortic Occlusion Device, REF: ICF100, 10.5 Fr (3.5 mm) X 39.4" (100 cm), Rx Only, Sterile EO The IntraClude intra-aortic occlusion device is indicated for use in patients undergoing cardiopulmonary bypass. The IntraClude intra-aortic occlusion device occludes and vents the ascending aorta when the balloon is inflated.

Why the product was recalled

Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the operative site may be obscured, and the device will need to be exchanged or operative strategy would need to change, including placement of an external cross-clamp, conversion to an open procedure, or performing the procedure under fibrillation.

Distribution information

Worldwide Distribution: US (nationwide) to states of: UT, GA, NJ, CA, OH, PA, AL, FL, TN, TX, ME, MA, IN, MS, NY, VA, MI, IL, CO, OR, WA, SC, AZ, SD, MO, CT, OK, NV, WV, WI, NC, MN, KY; and countries of: United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Czech Republic, Germany, France, United Kingdom, Italy, Netherlands, Poland, Romania, and Sweden.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.