FDA recall record

Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees recall — FDA Z-0197-2015

A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle…

Class IIclassification
Z-0197-2015official record ID
Nov 19, 2014report date

Recall record at a glance

Official recordZ-0197-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntegra LifeSciences Corp.
Report dateNov 19, 2014
Recall initiation dateOct 24, 2014
Quantity12 units
LocationPlainsboro, NJ, United States

Official product description

Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.

Why the product was recalled

A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

Distribution information

Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.