FDA recall record

Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) recall — FDA Z-0831-2015

Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was…

Class IIclassification
Z-0831-2015official record ID
Dec 24, 2014report date

Recall record at a glance

Official recordZ-0831-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntegra LifeSciences Corp.
Report dateDec 24, 2014
Recall initiation dateNov 7, 2014
Quantity2,506 units
LocationPlainsboro, NJ, United States

Official product description

Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA

Why the product was recalled

Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.

Distribution information

Nationwide, Foreign

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.