FDA recall record

Heartware Instructions for Use and Patient Manual for HeartWare HVAD System recall — FDA Z-1469-2021

IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that…

Class IIclassification
Z-1469-2021official record ID
Apr 28, 2021report date

Recall record at a glance

Official recordZ-1469-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmHeartware, Inc.
Report dateApr 28, 2021
Recall initiation dateFeb 26, 2021
Quantity23123 (11292 US)
LocationMiami Lakes, FL, United States

Official product description

Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.

Why the product was recalled

IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604

Distribution information

Worldwide distribution, including: US nationwide, Italy, Great Britain, Germany, Japan

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 13, 2023

Heartware PUMP 1103 HVAD recall — FDA Z-0478-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0482-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0481-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0475-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0479-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP MCS1705PU HVAD recall — FDA Z-0483-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.