Recall record at a glance
| Official record | Z-1438-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Heartware, Inc. |
| Report date | Apr 28, 2021 |
| Recall initiation date | Feb 26, 2021 |
| Quantity | 23123 (11292 US) |
| Location | Miami Lakes, FL, United States |
Official product description
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.
Why the product was recalled
IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible click", to clarify under what alarm conditions a beep will not be heard, and identify alternate confirmation methods (visual indicator) for successful connection. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87603
Distribution information
Worldwide distribution - US Nationwide distribution and the countries of Italy, Great Britain, Germany, Japan.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.