Recall record at a glance
| Official record | Z-1919-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | LivaNova USA |
| Report date | May 30, 2018 |
| Recall initiation date | Mar 13, 2018 |
| Quantity | 70 |
| Location | Arvada, CO, United States |
Official product description
Inspire 8 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050714 The intended uses for the two elements that constitute the oxygenator/reservoir integrated devices are: INSPIRE 8M Hollow Fiber Oxygenator: The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. The INSPIRE 8M is intended to be used for 6 hours or less. INSPIRE HVR (Hardshell Venous/Cardiotomy Reservoir): The INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, de-foams and filters venous blood and suction blood. INSPIRE HVR can be used post-operatively for chest drainage. The INSPIRE HVR is intended to be used for 6 hours or less.
Why the product was recalled
Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar code and readable code just beneath the barcode. Specifically, the expiration date is reported as DDMMYY instead of YYMMDD. The expiration date in the applicable field on the package labels are correct as printed.
Distribution information
US Distribution to the states of : NJ, CA, MN, MO, MA
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.