FDA recall record

Philips Ingenia Ambition X, Model No. 782109 recall — FDA Z-0810-2024

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the…

Class IIIclassification
Z-0810-2024official record ID
Jan 31, 2024report date

Recall record at a glance

Official recordZ-0810-2024
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusOngoing
Recalling firmPhilips North America
Report dateJan 31, 2024
Recall initiation dateDec 29, 2023
Quantity7 US; 16 ROW total
LocationCambridge, MA, United States

Official product description

Ingenia Ambition X, Model No. 782109

Why the product was recalled

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Distribution information

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.