FDA recall record

Toshiba American Medical Systems INFX-8000V Bi-Plane X-Ray Interventional System X-ray recall — FDA Z-0752-2016

When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter…

Class IIclassification
Z-0752-2016official record ID
Feb 17, 2016report date

Recall record at a glance

Official recordZ-0752-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmToshiba American Medical Systems Inc.
Report dateFeb 17, 2016
Recall initiation dateAug 14, 2015
Quantity4
LocationTustin, CA, United States

Official product description

INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems

Why the product was recalled

When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".

Distribution information

US Nationwide Distribution to OH and NY

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.