| Official record | Z-0355-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | B. Braun Medical, Inc. |
| Report date | Dec 13, 2023 |
| Recall initiation date | Nov 1, 2023 |
| Quantity | 33,820,976 units |
| Location | Allentown, PA, United States |
Official product description
Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SPACE PUMP IV SET; 362043 Priming volume: 39mL, length: 120in, SafeLine; 362050 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; 362420 SPACE PUMP SET 15 DROP W/2 SAFEDAY; 362432 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; 362911 Priming volume: 14mL, length: 84in, CARESITE, Spin-Lock; 363010 Priming volume: 21mL, length: 121in, SafeLine; 363019 Priming volume: 40mL, length: 123in, SafeLine; 363030 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 363032 20 drops/mL, priming volume: 20mL, length: 119in, SafeLine; 363230 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; 363410 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 363419 Priming volume: 41mL, length: 120in, CARESITE; 363420 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 363421 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 363422 Priming volume: 21mL, length: 130in, CARESITE; 363423 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; 363424 UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; 363430 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; 363433 UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 363901 60 drops/mL, priming volume: 11mL, length: 129in; 363902 Priming volume: 25mL, length: 130in; 363904 UNIV. 15 DROP PUMP SET, ASV; 470119 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 480254 SPACE PUMP IV SET 15D, 110 IN.; 480255 SPACE PUMP IV SET, 127 IN; 480257 60 DROP NITRO SPACE SET, 88 IN.; 480258 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 480263 SPACE PUMP IV SET 15D, 90 IN.; 480264 SPACE PUMP IV SET, 15 DRPS/ML, 95 IN.; 490037 Priming volume: 25mL, length: 136in; 490038 60 drops/mL, priming volume: 11mL, length: 129in; 490100 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490101 Priming volume: 19mL, length: 120in, CARESITE; 490102 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490103 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 490104 60 drops/mL, priming volume: 21mL, length: 115in, CARESITE; 490105 Priming volume: 39mL, length: 120in, CARESITE
Why the product was recalled
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.
Distribution information
Distribution US nationwide, Canada and Germany.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.