FDA recall record

Canon Medical System, USA Infinix-8000F with Catheterization Table CAT-850B or CAT-860B recall — FDA Z-2327-2020

Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Class IIclassification
Z-2327-2020official record ID
Jun 17, 2020report date

Recall record at a glance

Official recordZ-2327-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCanon Medical System, USA, Inc.
Report dateJun 17, 2020
Recall initiation dateMar 3, 2020
Quantity15
LocationTustin, CA, United States

Official product description

Infinix-8000F with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Why the product was recalled

Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Distribution information

US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.