Philips Incisive CT, Model Numbers 728144 and 728143 recall — FDA Z-0719-2024

If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the…

Class IIclassification
Z-0719-2024official record ID
Jan 24, 2024report date

Recall record at a glance

Official recordZ-0719-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPhilips North America
Report dateJan 24, 2024
Recall initiation dateDec 7, 2023
Quantity55 US; 1029 ROW
LocationCambridge, MA, United States

Official product description

Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871

Why the product was recalled

If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.

Distribution information

Domestic distribution to the following states: AL CA CO FL GA HI IL IN KS KY LA MD MI MO MS ND NE NJ NY OH OK PA RI SC TX UT VT WA International distribution worldwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.