FDA recall record

CooperSurgical INCA recall — FDA Z-3140-2026

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of…

Class Iclassification
Z-3140-2026official record ID
Sep 30, 2026report date

Recall record at a glance

Official recordZ-3140-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmCooperSurgical, Inc.
Report dateSep 30, 2026
Recall initiation dateAug 11, 2026
Quantity149 boxes (745 units)
LocationTrumbull, CT, United States

Official product description

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

Why the product was recalled

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

Distribution information

Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 30, 2026

CooperSurgical INCA recall — FDA Z-3139-2026

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the…

Class IFDA: OngoingCooperSurgical, Inc.
Medical DeviceJul 30, 2025

CooperSurgical HSG Procedure Tray recall — FDA Z-2135-2025

The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.

Class IIFDA: OngoingCooperSurgical, Inc.
Medical DeviceDec 25, 2024

K-Systems G85 Mini Incubator recall — FDA Z-0688-2025

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm…

Class IIFDA: OngoingCooperSurgical, Inc.
Medical DeviceDec 25, 2024

K-Systems T47 Warming Plate recall — FDA Z-0689-2025

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm…

Class IIFDA: OngoingCooperSurgical, Inc.