Recall record at a glance
| Official record | Z-0584-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | GN Hearing A/S |
| Report date | Dec 4, 2019 |
| Recall initiation date | Sep 18, 2019 |
| Quantity | 285 units |
| Location | Ballerup, Denmark |
Official product description
In-the-Ear hearing aids, Model Numbers: T9ITC-DW-LP, FT6ITC-D-MP, FT6ITE-D-MP, FT8CIC-MP, FT8ITC-W-HP, FT8ITC-W-LP, FT8ITC-W-MP, FT8ITE-DW-HP, FT8MIH-W-HP, FT8MIH-W-MP, LT5CIC-MP, LT5CIC-UP, LT5IIC-LP, LT5ITC-DW-MP, LT5ITC-DW-UP, LT5ITE-DW-HP, LT5ITE-DW-MP, LT5ITE-DW-UP, LT5MIH-W-HP, LT5MIH-W-MP, LT5MIH-W-UP, LT7CIC-HP, LT7CIC-LP, LT7CIC-MP, LT7CIC-W-MP, LT7IIC-LP, LT7ITC-DW-LP, LT7ITC-DW-MP, LT7ITC-MP, LT7ITC-W-HP, LT7ITC-W-MP, LT7ITE-DW-HP, LT7ITE-DW-MP, LT7ITE-DW-UP, LT7ITE-W-UP, LT7MIH-W-HP, LT7MIH-W-LP, LT7MIH-W-MP, LT9CIC-HP, LT9CIC-MP, LT9CIC-W-MP, LT9IIC-LP, LT9ITC-DW-HP, LT9ITC-DW-LP, LT9ITC-DW-MP, LT9ITE-DW-HP, LT9ITE-DW-MP, LT9ITE-DW-UP, LT9MIH-HP, LT9MIH-MP, LT9MIH-W-HP, LT9MIH-W-LP, LT9MIH-W-MP, TST17IIC-LP, TST17ITC-DW-HP, TST17ITE-HP, TST17MIH-W-UP, TST6CIC-HP, TST6IIC-LP, TST6ITC-MP, TST6ITC-W-HP, TST6ITC-W-MP, TST6ITE-DW-UP, TST6ITE-HP, TST6ITE-W-UP, TST6MIH-W-HP, TST9ITC-DW-MP, TST9ITE-DW-HP, TST9ITE-MP, TST9MIH-W-MP, VA8CIC-HP, VA8CIC-MP, VA8IIC-LP, VA8ITC-DW-HP, VA8ITC-DW-MP, VA8ITC-W-MP, VA8ITE-D-HP, VA8ITE-DW-HP, VA8ITE-DW-UP, VA8MIH-W-MP
Why the product was recalled
GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.
Distribution information
Nationwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Telephone.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.