FDA recall record

Implant Direct Sybron Manufacturing ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 recall — FDA Z-0185-2018

One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and…

Class IIclassification
Z-0185-2018official record ID
Dec 6, 2017report date

Recall record at a glance

Official recordZ-0185-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmImplant Direct Sybron Manufacturing, LLC
Report dateDec 6, 2017
Recall initiation dateOct 10, 2017
Quantity75 units
LocationWestlake Village, CA, United States

Official product description

ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Why the product was recalled

One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial. As a result, the cap could possibly not be screwed on properly and the sterilization compromised.

Distribution information

Worldwide distribution: US (nationwide) including Puerto Rico and country of Japan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.