Recall record at a glance
| Official record | Z-0980-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | Abiomed, Inc. |
| Report date | Feb 21, 2024 |
| Recall initiation date | Dec 27, 2023 |
| Quantity | 91,914 (65,857 US; 26,075 OUS) |
| Location | Danvers, MA, United States |
Official product description
Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 1000080 and 0048-0045 - Impella CP with SmartAssist ***Updated February 2024*** (10) 004413 - Impella 2.5 Set (12) 0046-0026 - Impella 5.0 Pump Set ROW (13) 0046-0037 - Impella 5.0 Pump Set APAC (15) 004680-AU - 5.0 Pump Set AU (16) 0048-0002 - Impella CP Pump Set, EU (17) 0048-0002-BR - Impella CP Pump Set BR (18) 0048-0004 - Impella CP Pump Set, Canada (19) 0048-0014 - Impella CP Smart Assist Set, EU (20) 0048-0024-JP - Impella CP Smart Assist Set, JP (21) 0048-0044 - Impella CP Smart Assist Set, Canada (22) 0048-0047 - Impella CP Smart Assist Set APAC (23) 005040 - Impella 2.5 IMC Pump Set EU (24) 005048-JP - Impella 2.5 Pump Set, Japan (25) 005060 - Impella 5.0 IMC Pump Set EU (26) 005064 - Impella 5.0 IMC Pump Set Canada (27) 005066-JP - Impella 5.0 Pump Set, Japan (28) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (29) 0550-0004 - Impella 5.5 with SmartAssist Set, CA (30) 1000115 - Impella CP Pump set, APAC (31) 1000211 - Impella 5.5 SmartAssist Set, JP (32) 1000302 - Impella CP with SmartAssist APAC (33) 1000361 - Impella 5.5 Set AU (34) 1000402 - Impella CP Smart Assist Set ***Updated March 15, 2024*** The following products were removed from the list of affected products because the Right-side devices are not in the scope of the warning to avoid left ventricle perforation: (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (11) 0046-0011 - Impella RP Pump Set, EU (14) 0046-0039 - Impella RP Set APAC
Why the product was recalled
IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
Distribution information
Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.