FDA recall record

Imactis Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # recall — FDA Z-1264-2021

Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping…

Class IIclassification
Z-1264-2021official record ID
Mar 24, 2021report date

Recall record at a glance

Official recordZ-1264-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmImactis
Report dateMar 24, 2021
Recall initiation dateFeb 11, 2021
Quantity105 kits
LocationSt Martin D Heres, N/A, France

Official product description

Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100

Why the product was recalled

Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory

Distribution information

NJ, NY, WI

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.