Recall record at a glance
| Official record | Z-2022-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class III |
| FDA record status | Ongoing |
| Recalling firm | Randox Laboratories Ltd. |
| Report date | Jul 5, 2023 |
| Recall initiation date | May 4, 2023 |
| Quantity | 30 kits |
| Location | Crumlin (North), N/A, Ireland |
Official product description
Immunoassay Premium Control Tri Level-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes Catalog Number: IA2633
Why the product was recalled
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
Distribution information
CA, IL, VA, WA, WI including PR
How to read the classification
FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.