FDA recall record

Medical Intelligence Medizintechnik Gmbh IGuide System, software versions: 2.2.0, 2.2.1 recall — FDA Z-2082-2019

Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly…

Class IIclassification
Z-2082-2019official record ID
Jul 31, 2019report date

Recall record at a glance

Official recordZ-2082-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedical Intelligence Medizintechnik Gmbh
Report dateJul 31, 2019
Recall initiation dateJul 18, 2019
Quantity496 units
LocationSchwabmunchen, N/A, Germany

Official product description

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

Why the product was recalled

Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.

Distribution information

Worldwide Distribution - US Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.