FDA recall record

Beckman Coulter iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem recall — FDA Z-1893-2018

A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result…

Class IIclassification
Z-1893-2018official record ID
May 23, 2018report date

Recall record at a glance

Official recordZ-1893-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateMay 23, 2018
Recall initiation dateDec 4, 2017
Quantity14340 units
LocationBrea, CA, United States

Official product description

iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to automate the urine chemistry analysis profile using iChem VELOCITY Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterase, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChem VELOCITY strips are intended for use with the iCHEM VELOCITY analyzer. In particular they are not intended for visual reading. The iChem VELOCITY test strips are not intended for visual reading. The iChem VELOCITY is not intended to be used as a Point of Care (POC) analyzer. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. Tests performed using the iCHEM VELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.

Why the product was recalled

A subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous results for leukocytes.

Distribution information

Worldwide Distribution and US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.