FDA recall record

DePuy Orthopaedics I.M. Hole Locator Instrument. Orthopedic manual surgical instrument. The recall — FDA Z-0908-2015

Use of excessive force when impacting may lead to intra-operative femoral fracture.

Class IIclassification
Z-0908-2015official record ID
Jan 14, 2015report date

Recall record at a glance

Official recordZ-0908-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDePuy Orthopaedics, Inc.
Report dateJan 14, 2015
Recall initiation dateNov 12, 2014
Quantity1130
LocationWarsaw, IN, United States

Official product description

I.M. Hole Locator Instrument. Orthopedic manual surgical instrument. The IM Hole locator guides the IM drill during IM Hole preparation. The instrument is positioned between the condyles and aligned with the anatomical axis. Pins can be placed in the IM hole locator to assist with additional stability. The drill bush may also be lightly impacted onto the distal femur for additional stability.

Why the product was recalled

Use of excessive force when impacting may lead to intra-operative femoral fracture.

Distribution information

Worldwide Distribution: US (nationwide) including states of: AZ, CA, CO, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, NC, NV, NY, OH, PA, TN, TX, VA, WA, and WI; and countries of: Canada, Australia, Austria, Belgium, Czech Republic, France, Germany, India, Israel, Italy, Japan, Malaysia, Russia, South Africa, Switzerland, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.