FDA recall record

Nitinol Devices and Components HydroFinity Hydrophilic guidewire;Nitinol Guidewire with recall — FDA Z-1819-2014

Reports of outer polymer jacket to the core wire being damaged during use on the HydroFinity Guidewire.

Class Iclassification
Z-1819-2014official record ID
Jul 2, 2014report date

Recall record at a glance

Official recordZ-1819-2014
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmNitinol Devices and Components, Inc.
Report dateJul 2, 2014
Recall initiation dateMay 28, 2014
Quantity8485 in US, 9595 - ROW - total, all models
LocationFremont, CA, United States

Official product description

HydroFinity Hydrophilic guidewire;Nitinol Guidewire with Hydrophilic Coating, Sterile. 0.035 OD, Stiff Shaft, Angled Tip, 180 CM Length; Manufactured for Covidien, Plymouth, MN; Manufactured by: NDC Inc., Fremont, CA Manufactured in Costa Rica. The HydroFinity Hydrophilic guidewire facilitates the introduction and placement of catheters and interventional devices to the desired anatomical location during diagnostic or interventional procedures.

Why the product was recalled

Reports of outer polymer jacket to the core wire being damaged during use on the HydroFinity Guidewire.

Distribution information

Worldwide Distribution - USA (nationwide) and Internationally to Italy, France, Norway, Sweden, Germany, Belgium, United Kingdom and Canada.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.