FDA recall record

DeRoyal Industries HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units recall — FDA Z-0345-2016

DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered…

Class IIclassification
Z-0345-2016official record ID
Dec 9, 2015report date

Recall record at a glance

Official recordZ-0345-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDeRoyal Industries Inc.
Report dateDec 9, 2015
Recall initiation dateOct 14, 2015
Quantity16 cases of 50 units (800 eaches)
LocationPowell, TN, United States

Official product description

HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.

Why the product was recalled

DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system.

Distribution information

Distributed in the states of VA, AL, FL, and CA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.