FDA recall record

Leibel-Flarsheim Company Hydra Vision DR 60/80 System, Models 700539 and 700540 recall — FDA Z-1753-2013

A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum…

Class IIclassification
Z-1753-2013official record ID
Jul 24, 2013report date

Recall record at a glance

Official recordZ-1753-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLeibel-Flarsheim Company LLC
Report dateJul 24, 2013
Recall initiation dateApr 25, 2013
Quantity12 units
LocationCincinnati, OH, United States

Official product description

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

Why the product was recalled

A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.

Distribution information

Nationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.