Hospira RECEPTAL LINERS: 1L(1000mL), 1.5L (1500mL), 2L (2000mL) 1) List Number 43056-01 recall — FDA Z-2594-2016

Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with…

Class IIclassification
Z-2594-2016official record ID
Aug 24, 2016report date

Recall record at a glance

Official recordZ-2594-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHospira Inc.
Report dateAug 24, 2016
Recall initiation dateSep 10, 2013
Quantity67,410,861 liners and 67,410,861 lids/canisters in total
LocationLake Forest, IL, United States

Official product description

Hospira RECEPTAL LINERS: 1L(1000mL), 1.5L (1500mL), 2L (2000mL) 1) List Number 43056-01; 2) List Number 43023-01; 3) List Number 43025-01; 4) List Number 43025-11; 5) List Number 43025-21; 6) List Number 43025-25; 7) List Number 43038-05; 8) List Number 43043-01; 9) List Number 43044-01; 10) List Number 43044-11; 11) List Number 43046-01; 12) List Number 43047-01; 13) List Number 43041-01; 14) List Number 43024-01; 15) List Number 43027-01; 16) List Number 43027-14; 17) List Number 43001-01; 18) List Number 43064-01; 19) List Number 43046-05; 20) List Number 43042-11; 21) List Number 43042-01; 22) List Number 43044-05; 23) List Number 43044-15

Why the product was recalled

Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.

Distribution information

*** US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY), District of Columbia, Puerto Rico; *** FOREIGN: Australia, Canada, Costa Rica, Singapore, South Africa

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 16, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0362-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 16, 2016

Hospira Symbiq One Channel Infuser recall — FDA Z-0361-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 9, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0354-2017

Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply…

Class IIFDA: TerminatedHospira Inc.