FDA recall record

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030 recall — FDA Z-1700-2014

Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off…

Class IIclassification
Z-1700-2014official record ID
Jun 11, 2014report date

Recall record at a glance

Official recordZ-1700-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHologic, Inc.
Report dateJun 11, 2014
Recall initiation dateMay 13, 2014
Quantity11 units
LocationDanbury, CT, United States

Official product description

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.

Why the product was recalled

Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

Distribution information

Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.