Recall record at a glance
| Official record | Z-1144-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Hologic, Inc. |
| Report date | Mar 1, 2023 |
| Recall initiation date | Jan 16, 2023 |
| Quantity | 797 units |
| Location | Danbury, CT, United States |
Official product description
Hologic 3Dimensions/Selenia Dimensions Models: RM-SDM-00001-3D SDM-00001-2D SDM-00001-3D SDM-05000-3D3 SDM-05000-2DC SDM-00001-M3D SDM-05000-3DC SDM-SYS-6000-3D SDM-05000-2D3 SDA-SYS-3000-2D SDM-00001 SDM-SYS-9000-2D SDM-00001-M2D 3DM-SYS-STD 3DM-SYS-STD-NS SDM-05000-2AC SDM-05000-2A2 SDA-SYS-3000-3D SDA-SYS-3000-3D-UPS SDM-SYS-6000-3D-HTC SDM-05000-2D2 SDM-05000-3D2 SDM-SYS-6000-2D SDM-SYS-9000-3D
Why the product was recalled
C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual
Distribution information
US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.