FDA recall record

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH recall — FDA Z-2392-2016

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure…

Class IIclassification
Z-2392-2016official record ID
Aug 17, 2016report date

Recall record at a glance

Official recordZ-2392-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHitachi Medical Systems America Inc.
Report dateAug 17, 2016
Recall initiation dateJun 30, 2015
Quantity33 units
LocationTwinsburg, OH, United States

Official product description

Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Why the product was recalled

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.

Distribution information

US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.