FDA recall record

Hitachi Oasis MRI System - C-Spine Coil recall — FDA Z-1096-2017

The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.

Class IIclassification
Z-1096-2017official record ID
Feb 1, 2017report date

Recall record at a glance

Official recordZ-1096-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHitachi Medical Systems America Inc.
Report dateFeb 1, 2017
Recall initiation dateJan 6, 2017
Quantity180 units
LocationTwinsburg, OH, United States

Official product description

Hitachi Oasis MRI System - C-Spine Coil

Why the product was recalled

The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.

Distribution information

USA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Parma VA Medical Center, Outpatient Clinic. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.