FDA recall record

Polymer Technology Systems Henry Schein CardioChek Starter Kit, Reference Number 4060 HS recall — FDA Z-0834-2020

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create…

Class IIclassification
Z-0834-2020official record ID
Jan 29, 2020report date

Recall record at a glance

Official recordZ-0834-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPolymer Technology Systems, Inc.
Report dateJan 29, 2020
Recall initiation dateDec 2, 2019
Quantity44,448
LocationIndianapolis, IN, United States

Official product description

Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036

Why the product was recalled

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Distribution information

The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.