Cincinnati Sub-Zero Products Hemotherm CE, model 400CE, Heater-Cooling Device recall — FDA Z-1825-2017

Improperly rated fuses were used in the manufacture of the device.

Class IIclassification
Z-1825-2017official record ID
Apr 26, 2017report date

Recall record at a glance

Official recordZ-1825-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCincinnati Sub-Zero Products Inc.
Report dateApr 26, 2017
Recall initiation dateMar 9, 2017
Quantity390 devices
LocationCincinnati, OH, United States

Official product description

Hemotherm CE, model 400CE, Heater-Cooling Device

Why the product was recalled

Improperly rated fuses were used in the manufacture of the device.

Distribution information

Worldwide distribution. The recalled product was distributed to the following states: AL, AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, and WI. The recalled product was distributed to the following countries: Mexico, Taiwan. There are no Canada accounts for this recall. There is one (1) Mexico account for this recall: Koolfer S.A. DE C.V., San Isidro 168, Col. Pedregal De Santa Ursula Coapa, Del. Coyoacan, CP. Mexico D. F. 04600. There is one (1) government account for this recall: WRNMMC -PAA, 8965 Brown Dr, Bethesda, MD 20889.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.