HeartWare Ventricular Assist Device (HVAD) System recall — FDA Z-0726-2022

Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power…

Class IIclassification
Z-0726-2022official record ID
Mar 16, 2022report date

Recall record at a glance

Official recordZ-0726-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmHeartware, Inc.
Report dateMar 16, 2022
Recall initiation dateJan 27, 2022
Quantity197,889 units
LocationMiami Lakes, FL, United States

Official product description

HeartWare Ventricular Assist Device (HVAD) System

Why the product was recalled

Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.

Distribution information

Distribution throughout the United States. OUS distribution in Argentina, Armenia, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea (South), Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritius, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 13, 2023

Heartware PUMP 1103 HVAD recall — FDA Z-0478-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0482-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0481-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware HVAD recall — FDA Z-0475-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP 1104 HVAD recall — FDA Z-0479-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.
Medical DeviceDec 13, 2023

Heartware PUMP MCS1705PU HVAD recall — FDA Z-0483-2024

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm…

Class IIFDA: OngoingHeartware, Inc.