FDA recall record

Philips Electronics HeartStart MRx Monitor/Defibrillator, Model No recall — FDA Z-2035-2018

The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component.…

Class IIclassification
Z-2035-2018official record ID
Jun 6, 2018report date

Recall record at a glance

Official recordZ-2035-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips Electronics North America Corporation
Report dateJun 6, 2018
Recall initiation dateMay 3, 2018
Quantity5
LocationAndover, MA, United States

Official product description

HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable) Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

Why the product was recalled

The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.

Distribution information

Worldwide Distribution to accounts in OH and NC. Foreign distribution to Lebanon.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.