FDA recall record

Abbott Laboratories HbA1c Calibrators LN 4P52-02 consists of: 4P52-02 Hemoglobin A1c 1 x 1.6 recall — FDA Z-0135-2015

HbA1c Calibrators LN 4P52-02, lot 45063UQ12 may contain a value sheet for a different HbA1c calibrator lot. If calibrator lot 45063UQ12 is used in the…

Class IIclassification
Z-0135-2015official record ID
Nov 5, 2014report date

Recall record at a glance

Official recordZ-0135-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAbbott Laboratories
Report dateNov 5, 2014
Recall initiation dateApr 30, 2014
Quantity183 calibrator kits
LocationAbbott Park, IL, United States

Official product description

HbA1c Calibrators LN 4P52-02 consists of: 4P52-02 Hemoglobin A1c 1 x 1.6 mL and 1 x 1.6 mL. A1c Calibrators (lyophilized) contain hemoglobin and glycated hemoglobin from human whole blood. Prior to lyophilization, the calibrator matrix is an MES-buffered solution. Preservative: Ofloxacin. For use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 8000 and c 4000 Systems.

Why the product was recalled

HbA1c Calibrators LN 4P52-02, lot 45063UQ12 may contain a value sheet for a different HbA1c calibrator lot. If calibrator lot 45063UQ12 is used in the calibration of the HbA1c assay using values from an incorrect calibrator value sheet then all results, including quality control and patient, may be falsely elevated.

Distribution information

Worldwide Distribution in the countries of Austria, Canada, Cyprus, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Italy, Kuwait, Malaysia, Norway, Poland, Romania, Russia, Saudi Arabia, Thailand, Turkey, and United Arab Emirates.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 19, 2025

Alinity hq Analyzer recall — FDA Z-0489-2026

Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1652-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1654-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1655-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1656-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories