FDA recall record

Hamilton Medical HAMILTON-T1 Ventilator with software versions 1.1.2 or lower recall — FDA Z-0741-2013

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the…

Class Iclassification
Z-0741-2013official record ID
Feb 6, 2013report date

Recall record at a glance

Official recordZ-0741-2013
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmHamilton Medical Inc.
Report dateFeb 6, 2013
Recall initiation dateDec 19, 2012
Quantity97 total ventilators; 69 of part number 161005 and 28 of part number 161006
LocationReno, NV, United States

Official product description

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Why the product was recalled

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

Distribution information

US Nationwide Distribution including the country of Switzerland.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.