Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177 recall — FDA Z-1963-2015

An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+)…

Class IIclassification
Z-1963-2015official record ID
Jul 8, 2015report date

Recall record at a glance

Official recordZ-1963-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHamilton Medical, Inc.
Report dateJul 8, 2015
Recall initiation dateJun 1, 2015
Quantity154 total - all part numbers
LocationReno, NV, United States

Official product description

Hamilton Medical Infant Flow Sensor, single use, (1.6m), Part number 260177. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.

Why the product was recalled

An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.

Distribution information

US Nationwide Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.