Recall record at a glance
| Official record | Z-0668-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Boston Scientific Corporation |
| Report date | Dec 18, 2019 |
| Recall initiation date | Nov 18, 2019 |
| Quantity | 179 |
| Location | Marlborough, MA, United States |
Official product description
Habib EUS RFA, 20mm electrode active tip, Material Number/UPN: 6700 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.
Why the product was recalled
During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.
Distribution information
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.