| Official record | Z-2380-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Olympus Corporation of the Americas |
| Report date | Sep 8, 2021 |
| Recall initiation date | Jun 9, 2021 |
| Quantity | 6 boxes (5 per/box)=30 units |
| Location | Center Valley, PA, United States |
Official product description
Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general open surgical procedures when used in conjunction with the Gyrus ACMI G400 workstation Ref: 970010PC
Why the product was recalled
Expired adhesive used as an electrical insulation barrier and as a secondary retention bond failed and could result in foreign body / biologic reaction within the patient and / or electrical continuity to patient causing tissue burn to critical structures.
Distribution information
US Nationwide distribution in the states of CA, GA, MN, TN.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.