FDA recall record

Vascular Solutions Guardian II Hemostasis Valve, models 8210 and 8211 recall — FDA Z-1275-2016

Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close…

Class Iclassification
Z-1275-2016official record ID
Apr 13, 2016report date

Recall record at a glance

Official recordZ-1275-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmVascular Solutions, Inc.
Report dateApr 13, 2016
Recall initiation dateMar 4, 2016
Quantity5283
LocationMaple Grove, MN, United States

Official product description

Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with diameters up to 8 F are inserted into the human vascular system. The devices have two seals: the low-pressure seal, and the high-pressure seal. Closure of the high-pressure seal, which is achieved when the nut can no longer rotate, secures the diagnostic/interventional device in position within the vasculature and allows for pressure injections up to 600 psi.

Why the product was recalled

Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.

Distribution information

Nationwide Distribution in the states of AL, AR, CA, CO, DE, FL, GA, ID, IL IN, IA,KY, LA, MD, MA, MI, MO, NE,NV, NH, NY, NC, OH, PA, SD, TN, TX, UT, VA, WA, and WV. OUS: Ireland

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.