FDA recall record

Terumo Medical Glidesheath Slender Hydrophilic Coated Introducer Sheath, 6FR - 10CM (0.021) recall — FDA Z-0215-2018

An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the…

Class IIclassification
Z-0215-2018official record ID
Dec 20, 2017report date

Recall record at a glance

Official recordZ-0215-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTerumo Medical Corp
Report dateDec 20, 2017
Recall initiation dateJun 7, 2017
QuantityN/A
LocationSomerset, NJ, United States

Official product description

Glidesheath Slender Hydrophilic Coated Introducer Sheath, 6FR - 10CM (0.021), REF/Product Code RM*CS6F10NA, STERILE, Rx ONLY

Why the product was recalled

An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath.

Distribution information

Nationwide and Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.