FDA recall record

Steris GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform recall — FDA Z-0241-2023

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear…

Class IIclassification
Z-0241-2023official record ID
Nov 23, 2022report date

Recall record at a glance

Official recordZ-0241-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSteris Corporation
Report dateNov 23, 2022
Recall initiation dateOct 17, 2022
Quantity5 units
LocationMentor, OH, United States

Official product description

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001

Why the product was recalled

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result

Distribution information

US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 14, 2024

Steris Hookup recall — FDA Z-2538-2024

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an…

Class IIFDA: OngoingSteris Corporation